Click Image for Gallery
Product Details:
Minimum Order Quantity | 1 Piece |
---|---|
Brand | Systronics |
Wavelength Range | 340 to 999 nm |
Application | Industrial Use, Laboratory Use |
Model | 1203 |
Display | 4 Line 20 Ch. LCD |
Baseline Flatness | +/-0.005 Abs |
Detector | 2-Silicon Photodiode |
Photometric Range | +/-2.5 Abs |
Stray Light | Less than 0.3%T at 370 nm |
Wavelength Accuracy | +/-2 nm |
Readout | 1 nm per step |
Bandwidth | 5 nm |
Photometric Accuracy | +/-0.005 Abs at 1.0 Abs (with NIST filter) |
Repeatability | +/-0.005 Abs |
Resolution | 0.1%T, 0.001Abs |
Drift | +/-0.003 Abs / hour at 500 nm |
Scan Speed | Slow, Medium & Fast |
Grating | 600 lines/mm |
Source | Tungsten halogen lamp |
Sample Holder | One (10 mm) fixed for Reference & Sample cuvette holder |
Minimum Volume | 1 ml in 4 ml cuvette with cell riser |
Operating Mode | Single wavelength, Multi wavelength (Max 8), Spectrum scan and Time scan |
K Factor Range | 0.100 to 9999 with floating decimal point |
Conc Range | 0.100 to 9999 with floating decimal point |
Printer Port | Epson compatible 80 column Dot Matrix |
Power | 230V +/-10%, 50 Hz |
Dimension | 450(W)x360(D)x130(H) mm |
Weight | 9 Kg (Approx) |
Accessories | Two 10 mm glass cuvette |
Systronics Visible Double Beam Spectrophotometer-1203 is a versatile visible double beam scanning instrument, having all processing facility like single wavelength, multi wavelength, scanning etc. This instrument offers outstanding performance in their ability to provide VIS spectral analysis of wide ranging samples, including Organic & Inorganic compound, Power Sectors, Industrial applications and R & D.
Salient Features:
Measuring Mode:
Absorbance, %Transmittance and Concentration (with by factor and by standard curve fitting)
RS232 Port:
PC Link data transfer
Data Processing:
Peak pick/Valley picks (6 points) Point picks (8 point), First derivative, Data saving facility (Four user defined files for each mode) Test saving facility (Four user defined files for each mode)
Optional Accessories:
Salient Features:
- Double Beam Optics
- Online Measurements
- Auto Wavelength & Base Line Correction
- Save / Recall Facility for Data / Test
Measuring Mode:
Absorbance, %Transmittance and Concentration (with by factor and by standard curve fitting)
RS232 Port:
PC Link data transfer
Data Processing:
Peak pick/Valley picks (6 points) Point picks (8 point), First derivative, Data saving facility (Four user defined files for each mode) Test saving facility (Four user defined files for each mode)
Optional Accessories:
- PC link software
- Epson compatible D. M. Printer
- Multi cuvette holders for 20/40/50 mm rectangular cuvette (Factory Installed)
Additional Information: