Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
Gas Chromatography
Gas Chromatography
Product Details:
Sr. No.
Main component
Detailed Specifications
1.0Gas ChromatographyGC with FID, NPD and ECD Detector along with original licensed windows-based software and Split-Spitless capillary inlet along with cool on column Inlet with Auto injector. GC must be capable to accommodate at least four two detectors & two injectors in working conditions simultaneously. System should support retention time locking facility with EPC resolution 0f 0.001 psi.
1.1Column Oven·Capable of housing at least two columns; operating temperature range: Upto 450 ° or more.

·Temperature Programming Ramps: 20 or more.

·Temperature set point resolution 0.1-degree C or more.

1.2Pneumatics·System must have pneumatic Electronic Flow Control for all injectors & detectors.
1.3Inlets (Two)· Split/split less capillary port

·Temperature: 400 °C or more

·Fully EPC

· Split ratio: 6000: 1 or more

· Pressure setting range 0–100 psi

· Cool on column Inlet

·Temperature 400 °C or more

·Pressure setting range 0–100 psi

1.4Nitrogen Phosphorus Detector (NPD)·Minimum detectable level: 0.1 pg N/s and 0.05 pf P/s

·Linear dynamic range: >105

·Maximum temperature: 400 °C

·Data acquisition rate up to 300 Hz

1.5Flame Ionization detector (FID)

 

·Electronic pressure and flow control of all gases

·Maximum temperature: 400 °C or more

·Minimum Detectivity: 1.5 pg C/sec or less

·Data acquisition rate up to 300 Hz

·Linear dynamic range: 107

1.6Electron Capture Detector (ECD)·Electronic pressure and flow control of all gases

·Maximum temperature: 400 °C or more

·Minimum Detectivity: 3.8 fg/ml or less

·Data acquisition rate up to 50 Hz

·Linear dynamic range: 104

1.7Auto Sampler·It should be able to do liquid injection with 2 ml vials.

· 50 or more Vials capacity.

1.8Software
  • Windows based software for evaluation of data, control and operation of entire system.
  • System security to be provided by the software and not by the operating system
  • User accounts, user privileges and audit trails are maintained by the software and not by the operating system
  • Compliance with GMP/GLP ,Electronic signature & Electronic Record rules
  • Software should have its own built in report .
  • Data that has been reviewed and approved may be locked to prevent further manipulations

· Software should offer extensive custom calculation capabilities (Concentration, calibration curve, peak soothing, peak subtraction etc.) eliminating the need for external applications to meet the laboratory need.

2.0Capillary Columns· Column for pesticides analysis. – Qty 2

· 100-meter column for Fatty acid. – Qty 2

3.0Consumables·Vials and Caps- 500

·Liner- 10 each for split and split-less

·Ferrules- 10

·Column nut- 10

·Septa- 100

· Auto sampler syringe- 4

4.0Computer·Compatible PC with 8 GB RAM, 512 GB HDD and monitor.
4.1Printer· Laser Jet Printer
4.2Standards·FAME standards
5.0Gas Cylinders·Filled Gas Cylinders of Hydrogen, Nitrogen and Zero Air with Gas Purification Panel and double stage brass regulators.
5.1Delivery·Delivery till F.O.R Destination
5.2Instrument performance·Installation at our site with detailed demonstration of all performance specifications including repeatability, software features, do’s and don’ts etc.
6IQ/ OQ·IQ, OQ, PQ of instrument and Software should be provided along with document.
7On-site training·Site Training for 2-3 working days on operation and troubleshooting should be part of the installation and commissioning of the system
8Warranty·One year warranty from the date of installation.

·The vendor should have availability for ten years guaranteed parts and services.

·Current user’s / performance list in India of the quoted Model with contact details (Customer name, phone email id etc) and date of installation to be provided (Minimum 5 installations of the model quoted)and this will be taken very seriously while making the decision. Your post sales service feedback will be certainly a deciding factor.

 

9Compliance Statement
  • The supplier must submit technical brochures and proper application notes adequately explaining and confirming the availability of the features in the model of the equipment being quoted.
  • The supplier must submit a table indicating the compliance of the features of the model of the equipment being quoted with those given in the indent.
  • Features not matching – must be clearly indicated.
  • Additional features and features in the quoted equipment which are better than those in the indent – may be clearly explained.
  • Bids submitted without Technical specifications manuals of each module quoted will not be evaluated for technical compliance.

·         All pre-installation requirements must be attached

 