Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
ELLIPSE ACCESS ANALYZER

SYSTEM DESCRIPTION

Open, random access, counter-top instrument for chemical and immunoturbidimetric clinical analysis.


METHODOLOGIES

Kinetic, fixed time, end point, bichromatic, differential, unlimited different methods can be programme.


SAMPLE LOADING

Continuous loading on 4 separate racks, 8 positions each. STAT samples can be placed at any time in any of the 4 racks and be run immediately.


SAMPLE CONTAINERS

Primary tubes diameter from8 to 16mm; height up to 100mm. Cups 3 ml for minimum quantity of serum.


SYSTEM DESCRIPTION

Open, random access, counter-top instrument for chemical and immunoturbidimetric clinical analysis.


METHODOLOGIES

Kinetic, fixed time, end point, bichromatic, differential, unlimited different methods can be programmed.


OPERATIVE MODE

  • Profile
  • Batch
  • Selective
  • STAT


THROUGHPUT

Up to 138 tests/hour.


SAMPLE LOADING

  • Continuous loading on 4 separate racks, 8 positions each.
  • STAT samples can be placed at any time in any of the 4 racks and be run immediately


SAMPLE CONTAINERS

  • Primary tubes diameter from 8 to 16 mm; height up to 100 mm.
  • Cups from 1 ml to 3 ml


SAMPLING

  • Sample volume: 2.0 to 99 microliters (1 μ increments)
  • Reagent volume: 3.0 to 548 microliters (1 μ increments)
  • Capacitive level sensor
  • Reagent preheating to 37 °C
  • Automatic sample probe washing (internal and external)
  • Probe wash cycle is operator programmable


REAGENTS

24 positions available for reagent containers.


STANDARDS AND CONTROLS

Up to 12 positions.


MINIMUM REACTION VOLUME

200 microliters.


READING CUVETTES

120 semi – disposable, optic grade cuvettes divided among 6 racks (20 each).


REACTION TEMPERATURE

Air bath at 37°C + 0.2.


OPTIC MODULE

  • Double beam
  • Up to 9 interferential filters + 1 position available for an optional filter
  • Narrow band filters 340, 380, 405, 492, 510, 546, 578, 620, 690 (optional)
  • FWHM: 10nm + 2nm
  • 6V/10W halogen bulb
  • Optic path length: 6 mm
  • Resolution: 0,0005 Abs
  • Linearity: up to 4,100 Abs


OPTIONAL ACCESSORIES

Bar code reader (reagents and samples).

Write a review

Note: HTML is not translated!
    Bad           Good

ELLIPSE ACCESS ANALYZER

  • ₹0

Seller information

View profile
Please sign in to contact Allied Biotechnology India Pvt L