Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
LIASYS CINICAL CHEMISTRY RANDOM

Fully automatic, random access, continuous loading, benchtop analyzer for clinical chemistry and immunoturbidimetric assays.

ASSAY TYPES

End Point, Kinetic, Differential, Bichromatic, Fixed Time, Non-Linear. All methods are user programmable (up to 999 tests).

ON LINE REAGENTS

4 racks are provided to hold up to 33 reagents that are

Divided as follows:

  • 2 racks cooled by Peltier cells holding up to 20 reagents
  • 2 racks for 13 reagent positions and 14 standards and controls at room temperature
  • 2 additional positions for diluents


DESCRIPTION

Fully automatic, random access, continuous loading, benchtop analyzer for clinical chemistry and immunoturbidimetric assays.


ASSAY TYPES

End Point, Kinetic, Differential, Bichromatic, Fixed Time, Non-Linear. All methods are user programmable (up to 999 tests).


ON LINE REAGENTS

  • 4 racks are provided to hold up to 33 reagents that are divided as follows:
  • 2 racks cooled by Peltier cells holding up to 20 reagents
  • 2 racks for 13 reagent positions and 14 standards and controls at room temperature
  • 2 additional positions for diluents


SAMPLES LOADING

  • 5 racks for continuous loading of samples split in:
  • 64 positions for routine samples
  • 14 positions for STAT samples


PRIMARY TUBE SIZES

  • From 40 mm to 100 mm (height)
  • From 10 mm to 16 mm (diameter)


STAT SAMPLE PROCESSING

Allowed at any time.


THROUGHPUT

Up to 333 tests/hour with ISE (200 without ISE).


INCUBATION TEMPERATURE

37 °C.


READING SYSTEM

Direct photometry.


MINIMUM REACTION VOLUME

300 μl.


OPTIC SYSTEM

  • Photometer: multi-wavelength, dual channel optic
  • Wavelengths: up to 9 narrow band interferential filters 340, 380, 405, 492, 510, 546, 578, 620 (690 opt.)
  • Light Source: halogen lamp 6V/10W
  • Optical Path: 10 mm
  • Linear Range: 0.001 ÷ 2.500 Abs
  • Resolution: 0.0001 Abs


SAMPLING ARM

  • One mechanical arm performs all sampling operations with:
  • Capacitive liquid level sensing
  • Reagent pre-warming at 37 °C
  • Automatic probe washing


DILUTER

  • Dilutions are made without using traditional syringes with the following specifications:
  • Sample volume: 2.0 μl ÷ 99 μl (1 μl incr.)
  • Reagent volume: 3.0 μl ÷ 500 μl (1 μl incr.)


REACTION PLATE

  • Holds 60 semi-disposable cuvettes and perfoms the following operations:
  • Automatic cuvettes washing
  • Continuous monitoring of cuvettes quality
  • Incubation temperature: control at 37 °C ± 0.3


OPTIONS

  • Direct potentiometry ISE module for NA+, K+, Cl-
  • Positive barcode reader (samples)