Click Image for Gallery
Construction Details:
- Designed as per ICH guidelines to meet, WHO and USFDA requirement
- Fulfills storage conditions of 25° C - 60% RH, 40° C - 75% RH, 30° C - 65% RH, 25° C - 40% RH, 30° C - 35% RH
- Temperature Accuracy : ± 1.0° C
- Temperature Resolution : 0.1° C
- Humidity Range : 30% RH to 80% RH,
- Himidity Fluctuation : Maximum 3% RH
- Chamber lllumination : With 2 or 4 LED lights for lesser power consumption
- Compressor Watts : 1.5 KW
- Dry Heater Watts : 2.0 KW
- Wet Heater Watts : 3.0 KW
- Overall Power consumption : 2.5 KW Per phase (For 3 - phase connection)
- Power consumption : 7.0 KW ( For single phase connection)
- Capacitance type Humidity sensor enables direct display of RH and temperature
- (ASI - PLC) Application Specific Integrated Programmable Logic Controller with Touch screen HMI for precise control of temperature and humidity parameters - fully protected with password
- Colour graphic HMI System
- Auto change over to standby systems through PLC - Change over from main set to Redundant Set and Vice Versa
- Multipurpose Temperature and Humidity logger with printout facility
- In-built data logger for selected number of channels (4, 8, 12, 16, 32) Within the LCD colour Graphic controller
- ASIC system with HMI having capacity of inbuilt storage upto 2 GB - 38,00000 Storage readings (Nearly unlimited storage) - this is the controller storage capacity
- 21 CFR compliance software - part 11 USFDA duidelines
- 10 - 20 different Humidity chambers can be connected to EIE - 21 CFR Comliance software
- Highly ergonomic door handle with unique key lock
- Built up with 80 mm Puff Insulation for maximum thermal protection.
Specifications:
- Type:Stability Chamber
- Accuracy Grade: 0.5
- Display: Digital
- Weight: 500-1000Kg
- Power Source: AC220V