Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
Opti CCA TS Electrolyte Analyzers

Being a well-established organization, we bring forward an excellent quality range of Opti CCA TS Electrolyte Analyzers to our precious clients. The offered electrolyte analyzers are widely appreciated by our precious clients owing to their high functionality and long operational life. The provided electrolyte analyzers are precisely manufactured by our highly experienced professionals using the latest technology. Further, these electrolyte analyzers are available in the market at competitive prices.

Product Details:

  • OPTI CCA with Color Touch Screen - Fully Automatic Microprocessor controlled, point of Care Blood Gas, Electrolytes & Glucose Analyzer with standard accessories:
  • Measured Parameters:
  • pH, pCO2, pO2,
  • Na, K, Cl Ca,
  • THB , SaO2
  • Glucose, bun

Salient Features:

  • Storage of reagents at normal room temperature except bun (2-8°C)
  • Automatic sample aspiration
  • Portable (light weight and Mains cum battery operated with 8 Hrs back-up)
  • Electronic quality control
  • Barcode scanner for quality control information and can be conveniently entered using the on-board bar code scanner
  • Onboard thermal printer
  • Display of results on in-built active color touch screen
  • 200 Patient data
  • RS 232 Interface Infrared. ASCII
  • CE, UL, IVDD compliant
  • FDA approved
  • Barometric Pressure: 300 - 800 MM HG

Operating Parameters:

  • Minimum Sample Size: 125 UL
  • Sample Type: whole blood, serum, plasma
  • Sample Application: Syringe, capillary or Micro sampler
  • Sample Input: Automatic Aspiration
  • Analysis Time (Time to Result): < 120 sec.
  • Ambient temp: 10°C - 32 °C
  • Relative Humidity: 5% - 95% (Non condensing)

Calculated Parameters:

  • Actual bicarbonate (HCO3), Base Excess (BE)
  • Base Excess extra cellular fluid ,Base Excess Actual, Buffer basis (BB), Total CO2: (T CO2)
  • Standard bicarbonate, standard pH(STP, pH)
  • Oxygen content (O2CT), Hematocrit (HCT)
  • Hydrogen Ion Concentration, Alveolar - Arterial: Oxygen difference (AaDO2), Anion gap(AG)
  • Standardized ionized Calcium (pH = 7.4), p50

Input Parameters:

  • Patient temp, Hemoglobin type, p50, FlO2, RQ
  • Operation ID. Patient ID No., Patient Sex
  • Mean Corpuscular Hemoglobin Concentration (MCHC),Sample type art / ven / mix, ven / cap / cord / CPB
  • User - defined field, 3 User- defined fields: Puncture site: left radial, right radial, left brachial, right brachial, left femoral, right femoral, cord, scalp

Data Management:

  • Print Out: Built-in Thermal printer
  • Interface: RS 232 C, ASCII or ASTM format
  • Quality Control Data & standard reference cassette report: 1 month of 3 levels stored onboard

Security:

  • Password: up to 2 levels – User and administrator
  • QC lock out: User definable, Number of levels / selectable time period, Operator ID.

Electrical And Gas Supply:

  • Voltage / Frequency: 100 – 250 VAC / 16 VDC, 2.8 A / 50 – 60 Hz.
  • Power Consumption (Maximum): 110 VA
  • Battery operation: Up to 8 Hrs. (Rechargeable battery pack)
  • Calibration Gas: Disposable small cylinder

Dimensions / Weight:

  • Height X Width X Depth: 124 MM X 362 MM X 230 MM / 4.7 Inch X 14.2 Inch X 9.1 Inch
  • Weight: 4 Kg (9 LBS) W / O battery 5 kg (11 LBS) with battery

Write a review

Note: HTML is not translated!
    Bad           Good

Opti CCA TS Electrolyte Analyzers

  • ₹0

Seller information

  • Bioteknika Healthcare Pvt Ltd

  • Bioteknika Healthcare Pvt Ltd
  • Ahmedabad, India
  • Business Type :Supplier
View profile
Please sign in to contact Bioteknika Healthcare Pvt Ltd

Tags: general lab equipment, others