Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
Microprocessor Flame Photometer

Feature:

  • Useful for both ppm and meq application.
  • Measurement of four elements with single aspiration.
  • Automatic filter selection.
  • Curve fitting software (up to 5 Standards)
  • Calibration stored in memory.
  • Re-standardization facility with only one standard.
  • User friendly menu driven software.
  • Help key available at each & every operating step.
  • 20 Character x 4 Line alphanumeric LCD display.
  • Measurement results stored in memory.
  • Measured results can be recalled on display & printed.
  • Measurement results can be printed (Individual, Batch-wiser & Date-wise).
  • RS 232 interface
  • Air Compressor with air regulator & in-built air filter.
  • Facility for storage of Date & Time of results.
  • Auto Ignition Facility

TECHNICAL SPECIFICATIONS

Model

11-1382

Range

Na: 0-100 ppm Upto 250 megil, 1: 100 dil.

 K: 0-100 ppm Upto 250 megil, 1: 100 dil.

Ca: 0-140 ppm Upto 250 meq/I, 1: 100 dil.

 Li: 10-100 ppm Upto 250 med/1, 1: 100 dil.

Sensitivity

Na: 0.1 ppm, K: 0.01 ppm, Ca: 0.5 ppm, Li: 10 ppm

Optical Management

Simultaneous analysis of 3 elements

Filter

Microprocessor controlled stepper motor with

Selection

0.5 KVA stabilizer

Units

ppm, mg and % age Selectable

Reproductivity

±1% FS ±1 Digits

Curve Fit Accuracy

±2% of FS

Display

Large LCD display 20x4 lines graph through Rs232

Aspiration

4-5 Seconds

Time Scale Range

0 to 100% Sensitivity for full scale

Flame System

LPG & Dry Oil free air

Detector

Photodiode

Calibration

Up to 5 point Calibration with curve fitting software

Gas Control

Adjustable with knob

Ignition

Auto Ignition  

System Fuel Gas

LPG

Burner

Stainless Steel

Dimensions

365x245x224 mm Approx. (LxBxH)

Printer

Provision for attachment of Dot - Matrix printer of HP Laser-Jet/Inkjet printer (non-win printers) with Centronics parallel interface

RS232 Interface

Rs232 Connector interface available (PC Software at extra Cost)

Air Compressor

With Built-in Regulator & Air filter to deliver stable & Moisture/oil free air supply r 1 kg/cm2

Air Supply

By Oil free mini Compressor unit with pressure regulator

Weight

Main Unit: 8 kg (Approx.), Compressor: 6.5 kg. (Approx.)

Power Supply

95 to 270V AC 50-60 Hz with inbuilt stabilizer

Blank Air Supply Combustion Gas

By oil controlled by precision regulator

LPG controlled by precision regulator

Power

230V ± 10?, 50Hz

Dimension

270W x 192H x 240D mm

Weight

6.25 kg. (Approx)

                        
Standard Accessories

Operation Manual

1 No.

Dust Cover

1 No.

Mains Lead

1 No.

Air Pipe, Drain Pipe

1 Meter each

Gas Connection tube

2 Meters

Chimney

1 No.

Capillary Tubes

1 Meter

Cleaning wire

1 Set

Hose clamps

4 Nos.

Gas Ignitor

1 No.















Write a review

Note: HTML is not translated!
    Bad           Good

Microprocessor Flame Photometer

  • ₹0

Seller information

  • Lab Junction

  • Lab Junction
  • Panchkula, India
  • Business Type :Manufacturer
View profile
Please sign in to contact Lab Junction

Tags: METERS