Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
Millikelvin Thermometers

Millikelvin Thermometers:

MKT

Model: MKT 10

  • Measuring uncertainty over the whole measuring range: 0.01 °C
  • For at-line measurements with a handheld instrument
  • Lightweight and portable

Model: MKT 50

  • Measuring uncertainty over the whole measuring range: 0.001 °C (1 mK)
  • Reference thermometer for all applications and branches
  • Two separate probes for comparison calibration
  • Sensors available for ITS 90, EN 60751, and ASTM E1137

Choose from two Millikelvin Thermometers, MKT 10 and MKT 50, for measurement in the lab, at-line, or for convenient mobile use. MKT 10 is a small, very lightweight, and easily portable handheld thermometer. It comes with one sensor and a carrying case for safe transport. With its measuring accuracy of 0.01 °C it is ideal for at-line measurements and quick measurements on site in several industries. Your MKT 50 Millikelvin Thermometer measures the temperature to the highest level of accuracy in combination with two highly precise Platinum Resistance Thermometers. It is by far the best-priced thermometer for applications which require a low measuring uncertainty of 0.01 °C to 0.001 °C. (Sensor needs to be ordered separately for MKT 50)


Key Features 

Highest accuracy

  • EN 60751 for industrial Platinum Resistance Thermometers (PRT)
  • Resolution: 0.0001 °C (0.1 mK) with MKT 50 and 0.001 °C (1 mK) with MKT 10
  • Measuring uncertainty over the whole measuring range: 0.001 °C (1 mK) with MKT 50 and 0.01 °C with MKT 10
  • ITS-90 for standard thermometers (SPRT: Standard Platinum Resistance Thermometer) with MKT 50
  • With MKT 50 comparison calibration (PRT) or fixed point calibration (SPRT) is possible.


Flexible power supply and communication

  • Battery-operated
  • Ethernet interface for easy communication with DMA M instruments and the SVM series of viscometers for calibration
  • The high precision thermometers can be used in the lab by connecting them to the mains or in the field as portable thermometers working from the battery.
  • MKT 50: mains adapter
  • MKT 10: Power over Ethernet (PoE) – ecologically friendly power supply for less use of batteries


Long-time experience

Anton Paar has decades of experience in density measurement which goes hand in hand with highly accurate temperature measurement since accurate temperature measurement is the basis for accurate density measurement.


Versatile applications

  • Ideal for different applications: from temperature measurement in the production process to different calibrations in the lab
  • Highly accurate comparison calibrations of all types of thermometers in liquid baths are possible with MKT 50.
  • MKT 50 is a reference thermometer for all applications and branches.
  • Calibration of standard thermometers at temperature fixed points according to ITS‑90 with MKT 50
  • MKT 10 provides maximum flexibility and at-line measurements with a handheld instrument.


Available sensors

  • Highly stable Pt 100 sensors with factory or DAkkS certificates
  • Flexible sensors for adjustment of DMA density meters and the SVM series of viscometers
  • MKT 50: Two types of working SPRTs are available for the highest accuracy (up to 1.4 mK).
  • MKT 50: for ITS‑90, EN 60751, and ASTM E1137 sensors


Technical specifications :

MKT 50MKT 10
Measuring range
Temperature−260 °C to +962 °C
(−436 °F to +1764 °F)
0 °C to 100 °C
(32 °F to 212 °F)
Resistance0 Ω to 440 Ω100 Ω to 140 Ω
Resolution
Temperature0.1 mK (Pt 100)1 mK
Resistance40 μΩ0.4 mΩ
Measurement uncertainty*
Temperature< 1 mK (Pt 100)< 10 mK
Resistance< 400 μΩ< 4 mΩ
Measuring sensorPt 100 or Pt 25.5
(EN 60751 or ITS‑90)
Pt 100
(EN 60751)
Measuring current0.41 mArms1 mADC
Internal reference resistor400 Ω (Vishay Precision Group, VHP 101)400 Ω (Vishay Precision Group, VHP 101)
Number of sensor inputs21
Sensor connection4‑wire (LEMO 1S304)4‑wire, soldered
Measuring time1.44 seconds (for both channels)1 second (for one channel)
Data outputsRS‑232D (optional USB**)
LAN (Ethernet, 10 Mbit)
LAN (PoE, 10 Mbit)
Ambient operating temperature0 °C to 35 °C
(for highest accuracy 20 °C to 25 °C recommended)
0 °C to 35 °C
(for highest accuracy 20 °C to 25 °C recommended)
Power supplyPower adapter 7.5 V or
2× AA batteries
4× AA batteries or
Power over Ethernet (PoE)
Dimensions (L × W × H)240 mm × 190 mm × 110 mm
(9.5 in × 7.5 in × 4.3 in)
200 mm × 100 mm × 35 mm
(7.8 in × 3.4 in × 1.4 in)
Weightapprox. 2 kg (4.4 lbs)approx. 500 g (1.1 lbs)

*) refers to the calibration of the reference resistor without contribution of the sensor (confidence level: 95 %, number of measurement values: 50, reference temperature: 23 °C)

**) with USB/RS‑232 adapter