Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
Domed Cooling Stage for Four-Circle Goniometers

Domed Cooling Stage for Four-Circle Goniometers:

DCS-500

  • For in-situ X-ray diffraction studies at low and high temperatures on four-circle goniometers and XYZ stages
  • Temperature range from -180 °C to +500 °C
  • For the investigation of layered structures, phase transformations, and stress in crystalline material


The DCS 500 Domed Cooling Stage is the smallest attachment for in-situ X-ray diffraction studies at low and high temperatures on four-circle goniometers and XYZ stages. With a temperature range from -180 °C to +500 °C and a theta angle from almost 0° to 90° 2Θ for all Φangles, DCS 500 is the ideal sample stage for in-situ studies of layered structures and for investigating phase transformations and stress in crystalline material.


Key Features

Exact sample temperature and variable sample environment

  • Liquid nitrogen cryostat for fast cooling down to -180 °C
  • Resistance heater for fast heating up to +500 °C
  • Accurate PT100 RTD close to the sample for precise temperature measurement and control
  • Highly X-ray transparent graphite dome for measurements under vacuum, in air and other gases

Compact design

  • Smart and variable sample fixing allows reflectivity measurements
  • Low weight (850 g)
  • Thin, vacuum-insulated liquid nitrogen hoses with high flexibility down to -180 °C


Technical specifications :

Operating temperature
in air, inert gas

25 °C to 500 °C
In vacuum (< 5x10-1 mbar)-180 °C to 500 °C
Temperature measurementPT100
AtmospheresVacuum (10-1 mbar), air, inert gas
Max. operating pressure0.3 bar above atmospheric pressure
Sample holder materialNickel-plated copper alloy
Sample diametermax. 25 mm
Dome materialGraphite
TransmissionCu Kα 72 %
Measuring geometryReflection