Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
Diet Beverage Analyzing system

Analyzing system:

Diet Beverage

  • Determines the %Diet concentration in colored, uncolored, and turbid diet drinks
  • Ready to analyze regular soft drinks next to diet drinks
  • Results in 3 to 5 minutes
  • Optional pH measurement

Anton Paar’s Diet Beverage Analyzing System is the ideal solution for your diet drink. The system determines the %Diet concentration in diluted diet syrups and degassed final beverages. Combine the master instrument with Anton Paar’s diet module for your daily beverage quality control in the production process. For automated filling and/or cleaning, the Diet Beverage Analyzing System can be upgraded with different types of sample changers. Benefit from a unique solution for your individual requirements.


Analyze all kinds of diet syrups and degassed beverages

Simply combine DMA 4500 M with a DietQC ME for a detailed diet beverage analysis to get the full range of vital parameters. Your options are endless: You can also choose the most accurate density meter in the world, DMA 5000 M, to measure turbid samples, DietQC ME with the option for uncolored drinks, or a module for pH determination. The Diet Beverage Analyzing System is also applicable for sugared soft drinks.

With only one measuring system, you can cover your whole soft drink portfolio.


Be ready for future requirements in beverage quality control

With the Diet Beverage Analyzing System, you can easily automate your measurements. Anton Paar’s sample changers free you from spending too much time on standard tasks: Choose Xsample 320 for automated filling of a single sample and Xsample 520 to fill up to 96 samples in a row. Xsample 330 not only fills the sample automatically, it also performs a cleaning routine right after the measurement to keep the instrument measuring accurately over a long time.

In case you want to get results for CO2 in your final product, Anton Paar also has the right solution for you: Simply upgrade to the Packaged Beverage Analyzer for soft drinks. 


The most accurate results for the analysis of diet beverages

Make the most out of your soft drinks and reach maximum efficiency in your production. When measuring diet beverages, DietQC ME with integrated absorption measurement guarantees reproducible results with a minimum of reference measurements. The Diet Beverage Analyzing System with an integrated DMA 5000 M provides highly accurate density values, based on the patented Pulsed Excitation Method, and is the high-end solution for turbid samples. 


Maximum level of usability

DietQC ME can be connected to the master instrument (either DMA 4500 M, DMA 5000 M, or Soft Drink Analyzer M, depending on your requirements) which has a number of intelligent features to make your daily life in the lab easier: FillingCheck™ confirms the bubble-free filling of the density and sound velocity measuring cell. If there is a problem with the filling, FillingCheck™ generates and records filling warnings. The state-of-the-art 10.4” PCAP touchscreen is robust yet sensitive. Measurement reports can be printed out in a standardized layout or the data can be transferred to your individual data system for full traceability of the measurements.


Technical specifications :
Measuring rangeRepeatability s.d.
Density (1)0 g/cm³ to 3 g/cm³DMA 5000 M/ SDA M: 0.000001 g/cm³
Sound velocity1000 m/s to 2000 m/s0.1 m/s
Concentration sugar0 °Brix to 15 °Brix<0.01 °Brix
Concentration sugar fresh/ inverted0 °Brix to 15 °Brix0.02 °Brix
Degree of inversion0 % to 100 %1 %

Diet concentration

For typical diet beverage compositions

0 % to 200 % or

0 mL to 100 mL NaOH or

0 g/L to 100 g/L TA

Colored diet drinks: 0.2 %

Uncolored diet drinks (via UV measurement): 0.4 %

Uncolored or turbid diet drinks (via density measurement): 0.4 %

pHpH 0 to pH 140.02 (in the range pH 3 to pH 7)

(1) according to ISO 5725

Measuring time3 to 5 minutes
Sample temperatureDMA M as master: 4 °C to 25 °C (39 °F to 77 °F)
SDA M as master: 15 °C to 25 °C (59 °F to 77 °F)
Sample volume per measurement30 mL
(approx. 20 mL per additional measuring module)
DimensionsDMA M/SDA M: 495 mm x 330 mm x 230 mm (19.5 in x 13 in x 9.1 in)
Xsample 520: 532 mm x 347 mm x 360 mm (20.9 in x 13.7 in x 14.2 in)
VoltageAC 100 to 240 V, 50/60 Hz
Power consumption190 VA (incl. external measuring modules)
Power inletAccording to IEC/EN 60320-1/C14, protection class I
FusesGlass tube fuses 5 mm x 20 mm; IEC60127-2; AC 250 V; T 5 AH
Environmental conditions (EN 61010)Indoor use only
Ambient temperature15 °C to 35 °C (59 °F to 95 °F)
Air humidity10 % to 90 % relative humidity, non-condensing
Pollution degree2
Overvoltage categoryII
Altitudemax. 3000 m (9800 ft)
Touchscreen10.4” TFT PCAP touchscreen 640 px x 480 px
Memory1000 measured data with/without camera pictures
Interfaces
  • 4x USB
  • Ethernet
  • CAN Bus
  • RS 232
  • VGA
RS232 printer settingsInterface: RS232; Baud rate: 9600; Parity: none; Stop bit: 1; Data bits: 8

Write a review

Note: HTML is not translated!
    Bad           Good

Diet Beverage Analyzing system

  • ₹0

Seller information

View profile
Please sign in to contact Anton Paar India Pvt. Ltd.

Tags: Analyzing System